The new Thorlabs [http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=4353 adjustable aspheres] (CFC-NX-A/B/C where N is the focal length) are a great improvement over the old models (CFC-N-A/B/C), but due to their shorter body length the SM1 adapter [http://www.thorlabs.com/thorProduct.cfm?partNumber=AD9.5F AD9.5F] does not leave the FC threads exposed.  Trying to attach a fiber like the one below will not work, because the female threads on the fiber cannot get inside the adapter to engate the male threads on the collimator.  This is not a problem with all fibers, but it is a problem with many fibers that we have.
+
The new Thorlabs [http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=4353 adjustable aspheres] (CFC-NX-A/B/C where N is the focal length) are a great improvement over the old models (CFC-N-A/B/C), but due to their shorter body length installing them in the the SM1 adapter [http://www.thorlabs.com/thorProduct.cfm?partNumber=AD9.5F AD9.5F] does not leave the FC threads exposed.  Trying to attach a fiber like the one below will not work, because the female threads on the fiber cannot get inside the adapter to engate the male threads on the collimator.  This is not a problem with all fibers, but it is a problem with many fibers that we have.
  
 
[[File:2010-03-03 17.54.31.jpg|400px]]
 
[[File:2010-03-03 17.54.31.jpg|400px]]

Revision as of 17:50, 7 March 2010

The new Thorlabs adjustable aspheres (CFC-NX-A/B/C where N is the focal length) are a great improvement over the old models (CFC-N-A/B/C), but due to their shorter body length installing them in the the SM1 adapter AD9.5F does not leave the FC threads exposed. Trying to attach a fiber like the one below will not work, because the female threads on the fiber cannot get inside the adapter to engate the male threads on the collimator. This is not a problem with all fibers, but it is a problem with many fibers that we have.

2010-03-03 17.54.31.jpg 2010-03-03 17.55.41.jpg

Using larger inner-diameter collimator adapters works, but then the collimator cannot be adjusted without removing it from the adapter.

One possible solution is to drill out the inner diameter of the AD9.5F so that the female threads can get inside the adapter.